Model | JERRY-I |
Size | 209 mm × 104 mm × 47.5 mm |
Net weight | 655 g |
Display | Color TFT - LCD / touch screen |
AC power adaptor | Input: 100 ~ 240 V AC, 50 / 60 Hz, 0.7 A |
Output: DC 5 V, 2 A | |
Battery | Li-ion rechargeable battery, 3.7 V / 7200 mAH |
Operating time | 15 hours with full battery capacity (condition: 25 ℃, NIBP working period is 15 minutes) |
Recharge time | 10 hours |
ECG | Input: 5- lead ECG cable, standard AAMI cable connector |
Lead selection: I, II, III, avR, avL, avF, C | |
Gain selection: × 1, × 2, × 4, × 8 | |
Sweep speed: 12.5, 25, and 50 mm/s | |
Frequency response: none filter: 0.05 ~ 100 Hz (- 3 dB) filter: 0.5 ~ 40 Hz (- 3 dB) | |
Calibration signal: 1 mV square wave | |
Heart Rate | Measuring range: 20 bpm ~ 300 bpm |
Renew time: 1 time per beat | |
Accuracy: ± 1 bpm | |
Alarm mode: audible and visual alarm | |
SpO2 | Measuring range: 0 % ~ 100 % |
Resolution: 1 % | |
Accuracy: ± 2 % (90 % ~ 99 %), ± 3 % (70 % ~ 89 %), unspecified (0 % ~ 69 %) | |
Pulse rate | Measuring and alarm range: 20 ~ 300 bpm |
Resolution: 1 bpm | |
Accuracy: ± 1 bpm | |
NIBP | Method: oscillometric |
Mode: manual, auto, stat | |
Measuring interval in auto mode: 1, 2, 3, 5, 10, 15, 20, 30, 45, 60 (min) | |
Measuring period in stat mode: 5 min | |
Cuff pressure range: 0 mmHg ~ 300 mmHg | |
Pulse rate range: 30 bpm ~ 250 bpm | |
Measurement range: Adult 30 mmHg ~ 255 mmHg Pediatric 30 mmHg ~ 160 mmHg | |
Over-pressure protection: Adult 300 mmHg Pediatric 220 mmHg | |
Pressure resolution: 1 mmHg | |
TEMP | Input: body surface thermal-sensitive resistor temperature sensor |
Measuring range: 0 ℃ ~ 50 ℃ | |
Accuracy: ± 0.1 ℃ | |
EtCO2 (optional) | Measuring Range: 0 ~ 150 mmHg |
Resolution: 1 mmHg | |
Accuracy: ± 2 mmHg @ < 5.0 % CO2 | |
Storage and review | Maximum data quantity: 3888 groups |
Maximum trend time: continuous 129.6 hours | |
Environment | Temperature: Operation: 0 ℃ ~ 45 ℃ Transportation and storage: - 25 ℃ ~ 85 ℃ |
Humidity: Operation: 30 % ~ 85 % Transportation and storage: 30 % ~ 95 % (non-condensing) |