Multi-para meter, 300F series
Product ID: MS28942
P/N: MS289421
With powerful features, intelligent system design and simple operation steps.
Model
PH300F
EC300F
M300F
Quantity
Please note:
unit ( MOQ:  1 unit )
1 unit/pack, 1 pack(s)/carton
Recommend related products
$ Inquiry after Sign-In

Model

PH300F

EC300F

M300F

Parameters

pH / Temp (mV)

EC / Temp (Resistivity / TDS / Sal)

pH / EC / ISE / DO / Temp (mV /

pX / Resistivity / TDS / Sal  / DO

Saturation)

pH

Range

-2.00 to 20.00 pH

/

-2.00 to 20.00 pH

Resolution

0.01 pH

/

0.1, 0.01 pH

Accuracy

± 0.01 pH

/

± 0.01 pH

Calibration Points

Up to 5

/

Up to 5

Standard Customization

Yes

/

Yes

Standard Recognition

NIST, GB, USA and DIN buffers

/

NIST, GB, USA and DIN buffers

mV

Range

-2000 to 2000 mV

/

-2000 to 2000 mV

Resolution

0.1 mV

/

0.1 mV

Accuracy

± 0.3 mV or ± 0.1 %

/

± 0.3 mV or ± 0.1 %

ISE

Range

/

/

1.000E-9 to 9.999E-9

Unit

/

/

mol / L, mmol / L, g / L, mg / L, μg / L, ppm, ppb

Resolution

/

/

Up to 4 significant digits

Accuracy

/

/

± 0.5 %

Calibration Points

/

/

Up to 5

pX

Range

/

/

-2.00 to 20.00 pX

Resolution

/

/

0.1, 0.01 pX

Accuracy

/

/

± 0.01 pX

Calibration Points

/

/

Up to 5

Conductivity

Range

/

0.000 μS / cm to 1000 mS / cm

0.000 μS / cm to 1000 mS / cm

Resolution

/

0.001 μS / cm minimum, changed with range

0.001 μS / cm minimum, changed with range

Accuracy

/

± 1.0% FS

± 1.0% FS

Reference Temperature

/

20, 25 ºC

20, 25 ºC

Calibration Points

/

Up to 3

Up to 3

Standard Recognition

/

84 μS / cm, 1413 μS / cm, 12.88 mS / cm

84 μS/cm, 1413 μS / cm, 12.88 mS / cm

Resistivity

Range

/

5.00 Ω·cm 20.00 MΩ·cm

5.00 Ω·cm 20.00 MΩ·cm

Resolution

/

0.01 Ω·cm minimum

0.01 Ω·cm minimum

Accuracy

/

± 1.0 % FS

± 1.0 % FS

TDS

Range

/

0.00 ppm 300 ppt

0.00 ppm 300 ppt

Resolution

/

0.01 ppm minimum, changed with range

0.01 ppm minimum, changed with range

Accuracy

/

± 1.0 %  FS

± 1.0 %  FS

Salinity

NaCl

Range

/

0.0 80.0 ppt

0.0 80.0 ppt

Resolution

/

0.1 ppt

0.1 ppt

Accuracy

/

± 2 ppt

± 2 ppt

Dissolved

Oxygen

Concentration

Sensor Type

/

 

Polarographic

Range

/

/

0.00 to 20.00 ppm

Resolution

/

/

0.01 ppm

Accuracy

/

/

± 0.10 ppm

Calibration Points

/

/

Air-saturated water or zero point

Barometric Compensation

/

/

Yes

Manual Salinity Factor Correction

/

/

Yes

Saturation

Range

/

/

0.0 to 200.0%

Resolution

/

/

0.1 %

Accuracy

/

/

± 2.0 %

Temperature

Range

-5 to 110 ºC, 23 to 230 ºF

Unit

ºC, ºF

Resolution

0.1 ºC

Accuracy

± 0.2 ºC

Measurement

 

Reading Mode

AutoRead( Fast, Medium, Slow), Timed, Continuous

Reading Prompts

Reading, Stable, Locked

Temp. Compensation

ATC, MTC

Data Management

Data storage

500 results each

GLP Features

Yes

Inputs

 

pH Electrode

BNC (Q9)

/

 BNC (Q9)

Temp./DO Probe

 

 

4-pin aviation connector

Temp. /EC Probe

/

5-pin aviation connector

 5-pin aviation connector

Reference Electrode

Banana

/

Banana

Outputs

USB

PC

RS232

Printer

Display Options

 

Backlight

Yes

Auto Shutdown

1 ~ 60 min, off

IP Rating

IP 54

Date and Time

Yes

General

 

Power 

AC Adapter, 100 - 240 V

AC input, DC 9 V output

Dimensions

242 x 195 x 68 mm

Weight

900 g

Multi-para meter, 300F series
Please select the sample parameters
Model
PH300F
EC300F
M300F