Model
|
CWM-301D
|
|
Dimension
|
1400 mm (H) × 850 mm (W) × 700 mm (D)
|
|
Weight
|
65 kg
|
|
Casters
|
Front wheels with lock
|
|
Screen
|
10.1 inch LCD touch screen
|
|
Display graphics
|
Waveforms of P-T, F-T, V-T
|
|
Power
|
AC 100 ~ 240 V, 50 Hz ± 2 %
|
|
Battery power
|
≥ 180 min
|
|
Gas configuration
|
O₂, N₂O, Air
|
|
Pipeline input range
|
280 ~ 600 kPa (40 ~ 85 psi)
|
|
Patient range
|
Adult / Pediatric
|
|
Modes of ventilation
|
VCV (Volume Control Ventilation)
|
|
A/C (Assist / control Ventilation)
|
|
SIMV (Synchronized Intermittent Mandatory Ventilation)
|
|
MANUAL (Manual)
|
|
PCV (Pressure Control Ventilation)
|
|
PSV (Pressure Support Ventilation)
|
|
STAND-BY
|
|
Ventilator parameter
|
Working type
|
Electronic control, pneumatic driven
|
|
Tidal volume range
|
20 ~ 1500 ml
|
|
Pressure range
|
50 ~ 60 cm H₂O
|
|
Respiration rate range
|
1 ~ 100 bpm (1 ~ 40 bpm under SIMV)
|
|
I:E range
|
4:1 ~ 1:8
|
|
Ptrigger range
|
- 20 ~ 20 cm H₂O (based on PEEP)
|
|
Flow trigger
|
0 ~ 15 L
|
|
Minute volume range
|
> 18 L/min
|
|
SIGH range
|
0 ~ 5 / 100
|
|
System standard
|
Hypoxic guard system
|
N₂O cut-off valve, O₂ concentration > 25 %
|
|
Safety valve
|
< 12.5 kPa
|
|
Fresh gas compensation
|
25 ~ 75 L/min
|
|
Flowmeter
|
5 tubes, cascade 0.1 ~ 15 L/min O₂, 0.1 ~ 15 L/min N₂O, 0.1 ~ 15 L/min Air
|
|
Volume of CO₂ absorber
|
2 L
|
|
Vaporizer
|
2 selectetac, maximum 2 mounts (Isoflurane / Enflurane / Sevoflurane / Halothane)
|
|
Monitoring
|
Display graphics
|
Waveforms of P-T, F-T, V-T, loops of P-V, V-F, F-P
|
|
Types
|
VTi, VTe, MV, Pmax, Pplateau, Ri, Cs, Freq, FiO₂ (optional), ETCO₂ (optional)
|
|
Alarm limits
|
Audio / Visual
|
No tidal volume, MV, Paw, FiO₂ limitation O₂, Air failure, AC power failure, Battery low
|
|
PEEP
|
Type
|
Intergrated, electronic control
|
|
Range
|
OFF, 3 ~ 30 cm H₂O
|
|
Pre-setting functions
|
Language
|
English
|
|
Self-test
|
Automatic
|
|
Optional
|
AGSS, Suction unit, Communication interface
|