Model |
Aquarius (SpO2 + NIBP + PR) |
|
Aquarius plus (ECG + SpO2 + NIBP + RESP + TEMP + PR) |
|
|
Dimension |
175 mm × 275 mm × 175 mm |
|
Weight |
< 2.5 kg under standard configuration |
|
Display |
8.4’’ TFT LCD screen |
|
Resolution: 600 × 800 |
|
|
Number of traces |
3, up to 7 ECG waveforms |
|
LAN |
1 standard RJ45 port |
|
WLAN |
IEEE 802.11 b/g/n |
|
USB |
2 USB interface |
|
ECG |
Lead type: 3-lead (I, II, III) / 5-lead (I, II, III, aVR, aVL, aVF, V) |
|
ECG waveform: 1 channel, 7 channels |
|
|
Display sensitivity: 2.5 mm/mV (× 0.25), 5.0 mm/mV (× 0.5), 10 mm/mV (× 1.0), 20 mm/mV (× 2.0) |
|
|
Wave sweep speed: 3.125 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s, 50 mm/s |
|
|
Band width: diagnostic mode: 0.05 Hz ~ 100 Hz monitor mode: 0.5 Hz ~ 40 Hz surgery mode: 1 Hz ~ 20 Hz strong filter mode: 5 Hz ~ 20 Hz |
|
|
CMRR > 100 dB |
|
|
Notch: 50 / 60 Hz notch filter can be set to on or off |
|
|
Differential input impedance > 5 MΩ |
|
|
Electrode polarization voltage range: ± 400 mV |
|
|
HR range: 15 ~ 350 bpm |
|
|
Baseline recovery time < 3 s after defibrillation (in monitor and surgery mode) |
|
|
Calibration signal: 1 mVp-p, accuracy ± 3 % |
|
|
RESP |
Measurement method: thoracic electrical bioimpedance |
|
Measuring lead: lead I, II |
|
|
Wave gain: × 0.25, × 0.5, × 1, × 2 |
|
|
Respiratory impedance range: 0.5 ~ 5 Ω |
|
|
0 ~ 150 bpm |
|
|
Baseline impedance: 500 ~ 4000 Ω |
|
|
Gain: 10 grades |
|
|
Scan speed: 3.125 mm/s, 6.25 mm/s, 12.5 mm/s, 25 mm/s |
|
|
NIBP |
Measurement method: automatic oscillometric method |
|
Operating mode: manual, automatic, continuous |
|
|
Measurement unit: mmHg / kPa selectable |
|
|
Typical measurement time: 20 ~ 40 s |
|
|
Measurement type: systolic, diastolic, mean |
|
|
Measurement range (mmHg): range of systolic pressure: adult 40 ~ 280, pediatric 40 ~ 200, neonatal 40 ~ 135 range of diastolic pressure: adult 10 ~ 210, pediatric 10 ~ 150, neonatal 10 ~ 95 range of mean pressure: adult 20 ~ 230, pediatric 20 ~ 165, neonatal 20 ~ 105 |
|
|
Maximum average error: ± 5 mmHg |
|
|
Maximum standard deviation: 8 mmHg |
|
|
Resolution: 1 mmHg |
|
|
Interval: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180, 240, 480 minutes |
|
|
Overpressure protection: software and hardware, double safety protection |
|
|
Cuff pressure range: 0 ~ 300 mmHg |
|
|
SpO2 |
Measurement range: 0 ~ 100 % |
|
Parameter monitoring: perfusion index (PI): 1 ~ 10, optional pleth variability index (PVI) |
|
|
Resolution: 1 % |
|
|
Accuracy: ± 2 % or ± 2 bpm |
|
|
Refreshing rate: 1 s |
|
|
Masimo SET SpO2 (optional) |
Measurement range: 0 ~ 100 % |
|
Resolution: 1 % |
|
|
Accuracy: ± 2 % (70 ~ 100 %, adult / pediatric, non-motion, low perfusion); |
|
|
Refreshing rate: 1 s |
|
|
TEMP |
Measurement method: thermistor |
|
Measuring range: 5 ~ 50 ℃ |
|
|
Resolution: 0.1 ℃ |
|
|
Measurement accuracy: ± 0.1 ℃ |
|
|
Pulse rate |
Range: 30 ~ 300 bpm |
|
Resolution: 1 bpm |
|
|
Accuracy: ± 2 bpm (non-motion), ± 5 bpm (motion) |
|
|
Refreshing rate: 1 s |
|
|
Infrared ear thermometer (optional) |
Displayed range: 34 ~ 42.2 ℃ |
|
Operation ambient temperature range: 10 ~ 40 ℃ |
|
|
Accuracy for displayed temperature range: ≥ 35 ℃ ~ ≤ 42.2 ℃ range ± 0.2 ℃, < 35 ℃ ~ ≥ 34 ℃ range ± 0.3 ℃ |
|
|
Masimo ISA sidestream CO2 (optional) |
Warm-up time: full accuracy within 10 seconds |
|
Sampling flow rate: 50 ml/min (+ / - 10 / min) |
|
|
Accuracy: 0 ~ 15 % (± 0.2 % of the reading), 15 ~ 25 % (unspecified) |
|
|
Measurement range: 0 ~ 25 % |
|
|
Rise time: 200 ms, typical at 50ml/min flow rate |
|
|
Total response time: within 3 seconds (with 2m nomoline sampling line) |
|
|
AWRR range: 0 ~ 150 bpm |
|
|
Masimo IRMA mainstream CO2 (optional) |
Measurement range: 0 ~ 25 % |
|
Warm-up time: full accuracy within 10 seconds |
|
|
Accuracy: 0 ~ 15 % (± 0.2 % of the reading), 15 ~ 25 % (unspecified) |
|
|
AWRR range: 0 ~ 150 bpm |
|
|
Recorder (optional) |
Built-in, thermal dot array |
|
Horizontal resolution: 16 dots/mm (25 mm/s paper speed) |
|
|
Vertical resolution: 8 dots/mm |
|
|
Paper speed: 12.5 mm/s, 25 mm/s, 50 mm/s |
|
|
Number of waveform channels: 3 |
|
|
Operation environment |
Input power: AC 100 ~ 250 V, 50 / 60 Hz |
|
Temperature: 5 ~ 40 ℃ |
|
|
Humidity: < 85 % |